• DocumentCode
    2453467
  • Title

    Development of the model for the optical monitoring of urea in spent dialysate

  • Author

    Tomson, R. ; Fridolin, I. ; Uhlin, F. ; Holmar, J. ; Lauri, K. ; Luman, M.

  • Author_Institution
    Dept. of Biomed. Eng., Tallinn Univ. of Technol., Tallinn, Estonia
  • fYear
    2012
  • fDate
    3-5 Oct. 2012
  • Firstpage
    179
  • Lastpage
    182
  • Abstract
    The aim of the study was to develop a mathematical model suitable for the estimation of urea concentration removed during dialysis through UV-absorbance. 29 haemodialysis patients on chronic three-times-a-week haemodialysis were studied in 6 separate studies. Double-beam spectrophotometer was used for the determination of UV-absorbance in the collected spent dialysate samples. A single wavelength (SW) and a multi-wavelength (MW) model were developed using stepwise regression utilizing urea values from the laboratory as the dependent parameter. For SW model R2 was 0.656 (calibration set) and 0.729 (validation set), for MW model R2 was 0.696 and 0.764, respectively. In summary, the urea concentration removed during dialysis can be estimated with UV-absorbance technique.
  • Keywords
    biomedical engineering; blood; calibration; patient monitoring; spectrophotometry; ultraviolet spectra; calibration set; double-beam spectrophotometry; haemodialysis patients; mathematical model; multiwavelength model; optical monitoring; single wavelength model; spent dialysate; ultraviolet absorbance; urea concentration estimation; Biomedical optical imaging; Blood; Calibration; Laboratories; Optical variables measurement; Predictive models; Wavelength measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electronics Conference (BEC), 2012 13th Biennial Baltic
  • Conference_Location
    Tallinn
  • ISSN
    1736-3705
  • Print_ISBN
    978-1-4673-2775-6
  • Type

    conf

  • DOI
    10.1109/BEC.2012.6376846
  • Filename