• DocumentCode
    2461118
  • Title

    Derivation of Scatter Kernel in CBCT Imaging System

  • Author

    Li, Heng ; Mohan, Radhe ; Zhu, X. Ronald

  • Author_Institution
    Dept. of Radiat. Phys., Univ. of Texas M. D. Anderson Cancer Center, Houston, TX
  • fYear
    2006
  • fDate
    Oct. 29 2006-Nov. 1 2006
  • Firstpage
    1851
  • Lastpage
    1854
  • Abstract
    An edge spread function method has been developed to determine the scatter kernel in kilo-voltage (kV) cone beam computed tomographic (CBCT) imaging system. We modeled the scatter as a depth dependent convolution kernel, and the kernel was derived for different object thickness up to 41 cm from measured data without assuming any empirical trial function for a Varian Trilogy CBCT system. The derived kernel can then be used for correcting the projection data, or incorporated in the reconstruction process.
  • Keywords
    computerised tomography; image reconstruction; CBCT imaging system; Varian Trilogy CBCT system; depth dependent convolution kernel; edge spread function method; kilo-voltage cone beam computed tomographic imaging; reconstruction process; scatter kernel derivation; Convolution; Deconvolution; Detectors; Electromagnetic scattering; Image quality; Image reconstruction; Kernel; Particle scattering; Thickness measurement; X-ray scattering;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Signals, Systems and Computers, 2006. ACSSC '06. Fortieth Asilomar Conference on
  • Conference_Location
    Pacific Grove, CA
  • ISSN
    1058-6393
  • Print_ISBN
    1-4244-0784-2
  • Electronic_ISBN
    1058-6393
  • Type

    conf

  • DOI
    10.1109/ACSSC.2006.355082
  • Filename
    4176892