Title :
A 50 ppm low current multi-stage clamp-on current transformer
Author :
So, E. ; Bennett, D.A.
Author_Institution :
Inst. for Nat. Meas. Stand., Nat. Res. Council of Canada, Ottawa, Ont., Canada
Abstract :
Summary form only given, as follows.An electronically-aided multi-stage technique is applied to a hand-held clamp-on current transformer which enables accurate low current measurements to be made over a wide dynamic range of operation, 0.5% to 100% of rated current, for use in conjunction with highly accurate power and energy meters, and harmonic power analyzers. The clamp-on current transformer can be used to measure currents in the range from 1 A to 200 A with ratio errors of less than 50 ppm in both magnitude and phase.
Keywords :
calibration; current transformers; electric current measurement; harmonic analysis; power meters; 1 to 200 A; dynamic range; electronically-aided multi-stage technique; energy meters; harmonic power analyzers; low current measurements; multi-stage clamp-on current transformer; power meters; ratio errors; Current measurement; Current transformers; Dynamic range; Energy measurement; Harmonic analysis; Instruments; Magnetic cores; Power system harmonics; Transformer cores; Watthour meters;