Abstract :
Two methods for heating fluids in microliter- to milliliter-scale reaction chambers in disposable bioassay cartridges are analyzed and compared. Inductive heating requires no electrical contact between the energy source and the cartridge and uses a very inexpensive component in the cartridge. Resistive heating with a surface mount component requires electrical interconnection, but is generally conducive to low-cost off-the-shelf components. Typical power consumption for both inductive heating and resistive heating is consistent with battery-powered operation. A finite element model for heating an injection-molded plastic cartridge with a surface-mount resistor has been developed and validated through experiments on a 40 mm × 10 mm × 7.5 mm injection molded polystyrene cartridge with embedded 1kΩ surface-mount resistors. A model of frequency-dependent heat generation in a novel inductive heating device is also presented.
Keywords :
Fluids; Inductors; Resistors; Temperature measurement; Water heating; Biological Assay; Computer-Aided Design; Electric Power Supplies; Equipment Design; Equipment Failure Analysis; Feedback; Heating; Miniaturization; Reproducibility of Results; Sensitivity and Specificity; Systems Integration;
Conference_Titel :
Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
Conference_Location :
Boston, MA
Print_ISBN :
978-1-4244-4121-1
Electronic_ISBN :
1557-170X
DOI :
10.1109/IEMBS.2011.6090085