Title :
Comparison of reproducibility of time-domain, spectral temporal mapping, and spectral turbulence analyses of signal-averaged electrocardiogram
Author :
Gautzner, J. ; Kulakowski, P. ; Hnatkova, K. ; Staunton, A. ; Malik, M.
Author_Institution :
Dept. of Cardiological Sci., St. George´´s Hospital Med. Sch., London, UK
fYear :
1993
fDate :
5-8 Sep 1993
Firstpage :
443
Lastpage :
446
Abstract :
A population of 28 healthy volunteers was examined in order to compare short- and long-term reproducibility of time domain, spectral temporal mapping and spectral turbulence analyses of signal averaged ECG. The study confirmed that the reproducibility of the time-domain analysis is high, especially when using 40 Hz high-pass filter. Similar results were obtained for frequency measured durations of total QRS complex and terminal low power region. Reproducibility of spectral turbulence analysis exploring the entire QRS complex gives also satisfactory results, but spectral turbulence analysis restricted to the low power region is less reproducible. The reproducibility of spectral temporal mapping which combines frequency domain with time measurements is poor
Keywords :
electrocardiography; medical signal processing; spectral analysis; time-domain analysis; 40 Hz; ECG analysis reproducibility; QRS complex; frequency domain measurements; healthy volunteers; high-pass filter; long-term reproducibility; short-term reproducibility; signal-averaged electrocardiogram; spectral temporal mapping; spectral turbulence analyses; terminal low power region; time-domain analysis; Electrocardiography; Filters; Frequency domain analysis; Frequency measurement; Power measurement; Reproducibility of results; Signal analysis; Signal mapping; Spectral analysis; Time domain analysis;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computers in Cardiology 1993, Proceedings.
Conference_Location :
London
Print_ISBN :
0-8186-5470-8
Type :
conf
DOI :
10.1109/CIC.1993.378408
Filename :
378408