DocumentCode
2469102
Title
Eye-surface conformable telemetric structure for polymer-based retinal prosthesis
Author
Jeong, Joonsoo ; Lee, Seung Woo ; Min, Kyousik ; Eom, Kyungsik ; Bae, So Hyun ; Kim, Sung June
Author_Institution
School of Electrical Engineering and Computer Science, Seoul National University, Seoul, Korea
fYear
2011
fDate
Aug. 30 2011-Sept. 3 2011
Firstpage
1097
Lastpage
1100
Abstract
In this research, inductive telemetric structure for retinal prosthesis was developed based on Liquid Crystal Polymers. For power and data transmission into the polymer-based retinal implant which is conformable to eye surface, the designed coil was thermo-formed into convex shape. The geometric parameters of the coil were optimized using finite element method (FEM) simulations for maximizing coupling coefficient and quality factor. The electrical properties of fabricated coils were characterized and their power and data transmission performance was tested. The properties of deformed structure were compared to those of the planar structure. In vivo experiment were also conducted to confirm the functionality of telemetry system in implanted conditions as well as to estimate the influence of biological media upon link properties.
Keywords
Coils; Implants; In vivo; Polymers; Prosthetics; Retina; Shape; Animals; Coated Materials, Biocompatible; Computer-Aided Design; Elastic Modulus; Equipment Failure Analysis; Humans; Polymers; Prosthesis Design; Reproducibility of Results; Sensitivity and Specificity; Telemetry; Visual Prosthesis;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
Conference_Location
Boston, MA
ISSN
1557-170X
Print_ISBN
978-1-4244-4121-1
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2011.6090256
Filename
6090256
Link To Document