• DocumentCode
    2470775
  • Title

    The design and fabrication of a closed loop steady flow system for the study of thermodilution

  • Author

    Humphrey, Charlene R. ; Cezeaux, Judy L. ; Schreiner, Steven

  • Author_Institution
    Mech. Eng. Dept., Western New England Coll., Springfield, MA, USA
  • fYear
    2002
  • fDate
    2002
  • Firstpage
    97
  • Lastpage
    98
  • Abstract
    A flow system modeling the right heart for the study of thermodilution was designed and constructed for use in an undergraduate teaching laboratory. The flow system was used to validate the thermodilution method for three known flow rates. Results show a linear relationship between the thermodilution and the known flow rates, however, thermodilution overestimates the true flow
  • Keywords
    biocontrol; biomedical education; biomedical equipment; biothermics; blood flow measurement; cardiology; closed loop systems; flow control; flowmeters; student experiments; temperature measurement; valves; ball valve; blood flow; cardiac output; closed loop steady flow system; computerized drive; constant head reservoir; flow system modeling; heat exchange; lower reservoir; pulmonary artery temperature; right atrium; right heart; thermistor; thermodilution; true flow; undergraduate teaching laboratory; variable area flowmeter; Arteries; Biomedical measurements; Blood; Catheters; Data acquisition; Fabrication; Head; Reservoirs; Temperature; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Bioengineering Conference, 2002. Proceedings of the IEEE 28th Annual Northeast
  • Conference_Location
    Philadelphia, PA
  • Print_ISBN
    0-7803-7419-3
  • Type

    conf

  • DOI
    10.1109/NEBC.2002.999483
  • Filename
    999483