• DocumentCode
    2474073
  • Title

    Evaluation of modal analysis techniques using physical models to detect osseointegration of implants in transfemoral amputees

  • Author

    Cairns, Nicola J. ; Adam, Clayton J. ; Pearcy, Mark J. ; Smeathers, James

  • Author_Institution
    Queensland Univ. of Technol., Brisbane, QLD, Australia
  • fYear
    2011
  • fDate
    Aug. 30 2011-Sept. 3 2011
  • Firstpage
    1600
  • Lastpage
    1603
  • Abstract
    Non-invasive vibration analysis has been used extensively to monitor the progression of dental implant healing and stabilization. It is now being considered as a method to monitor femoral implants in transfemoral amputees. This paper evaluates two modal analysis excitation methods and investigates their capabilities in detecting changes at the interface between the implant and the bone that occur during osseointegration. Excitation of bone-implant physical models with the electromagnetic shaker provided higher coherence values and a greater number of modes over the same frequency range when compared to the impact hammer. Differences were detected in the natural frequencies and fundamental mode shape of the model when the fit of the implant was altered in the bone. The ability to detect changes in the model dynamic properties demonstrates the potential of modal analysis in this application and warrants further investigation.
  • Keywords
    bone; dentistry; modal analysis; physiological models; prosthetics; bone-implant physical model; dental implant healing; dental implant stabilization; electromagnetic shaker; modal analysis technique; model dynamic properties; osseointegration; transfemoral amputees; Analytical models; Bones; Educational institutions; Implants; Modal analysis; Shape; Vibrations; Acceleration; Amputation Stumps; Biomimetics; Equipment Failure Analysis; Femur; Humans; Knee Prosthesis; Osseointegration; Prosthesis Design; Reproducibility of Results; Sensitivity and Specificity; Vibration;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
  • Conference_Location
    Boston, MA
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-4121-1
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2011.6090463
  • Filename
    6090463