11C-2 Minute Roughness Measurement Using Phase Tracking for Arterial Wall Diagnosis Non-Invasively In Vivo
Author
Cinthio, Magnus ; Hasegawa, Hideyuki ; Kanai, Hiroshi
Author_Institution
Lund Univ. Lund, Lund
fYear
2007
fDate
28-31 Oct. 2007
Firstpage
997
Lastpage
1000
Abstract
Early diagnosis of atherosclerosis is essential as many of the risk factors are life-style dependent. We suggest a new method to measure minute roughness of the size of micrometers of the arterial wall. The method was evaluated using three silicone phantoms sized 13 mum, 23 mum and 33 mum, respectively. The mean of the measured heights of the phantoms were 8.1 mum (SD 0.0), 23.3 mum (SD 0.2) and 29.6 mum (SD 0.1) in the forward direction, and 7.7 mum (SD 0.0), 21.9 mum (SD 0.2) and 27.3 mum (SD 3.6) in the backward direction, respectively. The phantom study shows very promising result and encourages to further evaluation and in vivo investigations.
Keywords
biomedical ultrasonics; phantoms; surface topography measurement; arterial wall diagnosis; atherosclerosis; phase tracking; roughness measurement; silicone phantoms; size 13 mum; size 23 mum; size 33 mum; Arteries; Atherosclerosis; Cardiovascular diseases; Electric variables measurement; Frequency; Imaging phantoms; In vivo; Phase measurement; Position measurement; Size measurement;