  • Abstract
    The feasibility of subharmonic contrast intravascular ultrasound (IVUS) imaging was investigated using a prototype nonlinear IVUS system and the commercial contrast agent DefinityTM. The system employed a mechanically scanned commercial catheter with a custom transducer element fabricated to have dual frequency peak sensitivity in the vicinity of 15 and 30 MHz. Experiments were conducted at a fundamental frequency of 30 MHz (F30; 25 % bandwidth) with on-axis pressures ranging from 0.12 to 0.79 MPa, as measured with needle a hydrophone. In vitro characterization experiments demonstrated the detection of 15 MHz subharmonic signals (SHI5) when pressure levels reached 360 kPa. The formation of SHI 5 images was shown, with tissue signals suppressed to near the noise floor and contrast to tissue ratios improved by up to 30 dB relative to F30, depending on the depth. In vivo experiments were performed using the atherosclerotic rabbit aorta model. Following the bolus injection of contrast agent upstream of the imaging catheter, agent was detected within the aorta, vena cava, and within the adventitia. Histology confirmed the presence of adventitial microvessels. These results provide a first in vivo demonstration of subharmonic contrast IVUS, and suggest its potential as a new technique for imaging the vasa vasorum.
  • Keywords
    biomechanics; biomedical ultrasonics; blood vessels; catheters; Definity contrast agent; IVUS imaging; adventitial microvessel; atherosclerotic rabbit aorta model; contrast agent bolus injection; frequency 15 MHz to 30 MHz; imaging catheter; mechanically scanned catheter; needle hydrophone; nonlinear IVUS system; pressure 0.12 MPa to 0.79 MPa; subharmonic contrast intravascular ultrasound; subharmonic signal detection; transducer element; vasa vasorum imaging; vena cava; Bandwidth; Biomedical imaging; Catheters; Frequency measurement; In vivo; Needles; Prototypes; Sonar equipment; Ultrasonic imaging; Ultrasonic transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2007. IEEE
  • Conference_Location
    New York, NY
  • ISSN
    1051-0117
  • Print_ISBN
    978-1-4244-1384-3
  • Electronic_ISBN
    1051-0117
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2007.269
  • Filename
    4409842