  • Abstract
    In the past, the optometry instrument calibrated in diopter and astigmatism were usually performed by using human eyes. However, the larger uncertainty of measurement will be induced while observing the measuring patterns using human eyes. Therefore, based on the ANSI/ISEA Z87.1-2010 standard, this study presents the development of automatic optometry system, which comprises of the machine vision, image analysis and processing, and automatic focus technique without human operation. In addition, the astigmatism measurement can be well completed in the electric moving stage. The experimental results show the automatic optometry system has the high accuracy and good repeatability.
  • Keywords
    calibration; computer vision; image processing; measurement uncertainty; vision defects; ANSI-ISEA Z87.1 standard; astigmatism measurement; automatic focus technique; automatic optometry instrument; calibration; human eyes; image analysis; image processing; machine vision; measurement uncertainty; Calibration; Focusing; Humans; Instruments; Lenses; Standards; Testing; automatic focus; image processing; optometry;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference (I2MTC), 2012 IEEE International
  • Conference_Location
    Graz
  • ISSN
    1091-5281
  • Print_ISBN
    978-1-4577-1773-4
  • Type

    conf

  • DOI
    10.1109/I2MTC.2012.6229156
  • Filename
    6229156