Measurement of respiration rate in preterm infants by laser Doppler vibrometry
Author :
Scalise, Lorenzo ; Ercoli, Ilaria ; Marchionni, Paolo ; Tomasini, Enrico Primo
Author_Institution :
Dipt. di Meccanica, Univ. Politec. delle Marche, Ancona, Italy
fYear :
2011
fDate :
30-31 May 2011
Firstpage :
657
Lastpage :
661
Abstract :
The continuous monitoring of patients recovered in intensive care units is a mandatory tasks for clinicians in order to check conditions and effects of therapies. Among the many signals and quantities observable, the ones related to the heart and particularly the lung´s activity are of primary interest. Today the majority of the measurement methods used for the assessment of the respiration rate are based on invasive methods (spriometry) which are often included into mechanical ventilator to which the patients, in need of assisted breathing, are attached. In this paper an optical, non-contact measurement method for monitoring of the respiration rate in pre-term infants is reported. It is based on the assessment of abdominal wall movements detected by a laser Doppler vibrometer and can be operated at a distance from the patient and outside the incubator. A total of 9 preterm patients have been monitored; respiration rates measured by a reference instrument (BabyLog 8000 Draeger plus) and the proposed non-contact method have been measured for a period of time of 60s. Data have been compared using the Bland-Altman analysis and results demonstrate an optimal correlation (R=0.97) among data and an uncertainty of the proposed method of ± 0.13s (K=2).
Keywords :
laser applications in medicine; paediatrics; patient monitoring; pneumodynamics; vibration measurement; BabyLog 8000 Draeger plus; Bland-Altman analysis; abdominal wall movement; continuous patient monitoring; intensive care unit; laser Doppler vibrometry; mechanical ventilator; preterm infant; respiration rate measurement; Instruments; Measurement by laser beam; Monitoring; Optical fibers; Optical sensors; Pediatrics; Skin; laser doppler vibrometry; non contact; respiration rate;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Medical Measurements and Applications Proceedings (MeMeA), 2011 IEEE International Workshop on
Conference_Location :
Bari
Print_ISBN :
978-1-4244-9336-4
Type :
conf
DOI :
10.1109/MeMeA.2011.5966740
Filename :
5966740
Link To Document :
بازگشت