DocumentCode
2488218
Title
Joint angle-based EMG amplitude calibration
Author
Hashemi, Javad ; Morin, Evelyn ; Mousavi, Parvin ; Hashtrudi-Zaad, Keyvan
Author_Institution
Dept. of Electr. & Comput. Eng., Queen´´s Univ., Kingston, ON, Canada
fYear
2011
fDate
Aug. 30 2011-Sept. 3 2011
Firstpage
4439
Lastpage
4442
Abstract
A calibration method is proposed to compensate for the changes in the surface electromyogram (SEMG) amplitude level of the biceps brachii at different joint angles due to the movement of the muscle bulk under the EMG electrodes for a constant force level. To this end, an experiment was designed, and SEMG and force measurements were collected from 5 subjects. The fast orthogonal search (FOS) method was used to find a mapping between SEMG from the biceps and force recorded at the wrist. Comparison between evaluation values from models trained with calibrated and non-calibrated SEMG signals revealed a statistically significant superiority of models trained with the calibrated SEMG.
Keywords
calibration; electromyography; force measurement; EMG amplitude calibration; SEMG amplitude; biceps brachii; fast orthogonal search; force measurement; joint angle; muscle movement; surface electromyogram; Calibration; Electrodes; Electromyography; Force; Joints; Muscles; Wrist; Adult; Calibration; Electromyography; Humans; Male; Muscle, Skeletal;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
Conference_Location
Boston, MA
ISSN
1557-170X
Print_ISBN
978-1-4244-4121-1
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2011.6091101
Filename
6091101
Link To Document