DocumentCode
2492674
Title
Conductive rubber belt to monitor respiratory changes
Author
Kim, Kyung-Ah ; Lee, In-Kwang ; Choi, Seong-Su ; Lee, Tae-Soo ; Cha, Eun-Jong
Author_Institution
Chungbuk Nat. Univ., Cheong-Ju
fYear
2006
fDate
22-25 Oct. 2006
Firstpage
74
Lastpage
76
Abstract
Conductive rubber material was molded in a belt shape to incorporate with patient´s pants for continuous monitoring of breathing activity. Dimension changes of the abdomen resulted in corresponding resistance changes, which successfully monitored respiration. When the standard air flow rate was simultaneously measured by a pneumotachometer, mean relative error was approximately 15%, presumably due to lack of chest signal. Wireless Zigbee transmission also provided a great convenience for monitoring. The present device would be of advantages for wearable applications for its simplicity, low cost, and being washable.
Keywords
biomedical equipment; biomedical measurement; conducting polymers; patient monitoring; pneumodynamics; rubber products; abdomen dimension changes; breathing activity monitoring; conductive rubber belt; pneumotachometer; resistance changes; respiratory changes; wireless Zigbee transmission; Abdomen; Belts; Conducting materials; Electrical resistance measurement; Fluid flow measurement; Measurement standards; Patient monitoring; Rubber; Shape; ZigBee;
fLanguage
English
Publisher
ieee
Conference_Titel
Sensors, 2006. 5th IEEE Conference on
Conference_Location
Daegu
ISSN
1930-0395
Print_ISBN
1-4244-0375-8
Electronic_ISBN
1930-0395
Type
conf
DOI
10.1109/ICSENS.2007.355721
Filename
4178559
Link To Document