• DocumentCode
    2492674
  • Title

    Conductive rubber belt to monitor respiratory changes

  • Author

    Kim, Kyung-Ah ; Lee, In-Kwang ; Choi, Seong-Su ; Lee, Tae-Soo ; Cha, Eun-Jong

  • Author_Institution
    Chungbuk Nat. Univ., Cheong-Ju
  • fYear
    2006
  • fDate
    22-25 Oct. 2006
  • Firstpage
    74
  • Lastpage
    76
  • Abstract
    Conductive rubber material was molded in a belt shape to incorporate with patient´s pants for continuous monitoring of breathing activity. Dimension changes of the abdomen resulted in corresponding resistance changes, which successfully monitored respiration. When the standard air flow rate was simultaneously measured by a pneumotachometer, mean relative error was approximately 15%, presumably due to lack of chest signal. Wireless Zigbee transmission also provided a great convenience for monitoring. The present device would be of advantages for wearable applications for its simplicity, low cost, and being washable.
  • Keywords
    biomedical equipment; biomedical measurement; conducting polymers; patient monitoring; pneumodynamics; rubber products; abdomen dimension changes; breathing activity monitoring; conductive rubber belt; pneumotachometer; resistance changes; respiratory changes; wireless Zigbee transmission; Abdomen; Belts; Conducting materials; Electrical resistance measurement; Fluid flow measurement; Measurement standards; Patient monitoring; Rubber; Shape; ZigBee;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensors, 2006. 5th IEEE Conference on
  • Conference_Location
    Daegu
  • ISSN
    1930-0395
  • Print_ISBN
    1-4244-0375-8
  • Electronic_ISBN
    1930-0395
  • Type

    conf

  • DOI
    10.1109/ICSENS.2007.355721
  • Filename
    4178559