    A New Cuff Unit for Measuring Instantaneous Blood Pressure at the Finger Artery by Local Pressurization

  • Author

    Song, Yilin ; Gao, Shumei ; Ikarashi, Akira ; Yamakoshi, Ken-ichi

  • Author_Institution
    Sch. of Mech. & Electr. Eng., Heilongjiang Univ., Harbin, China
  • fYear
    2009
  • fDate
    11-13 June 2009
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    A new cuff unit for measuring instantaneous blood pressure at the finger based on the volume-compensation method by local pressurization is developed, and its characteristics are evaluated. Results of measurement indicated that, the finger artery could be completely compressed using the local pressurization cuff, and the cuff-pressure controller is suitable for BP measurement in human. Also, the comparative measurements between the present cuff unit and the conventional band-type cuff indicated that, the venous congestion from the present unit is greatly reduced. Therefore, the present cuff unit is capable of non-invasively measuring instantaneous blood pressure at the finger artery in long term.
  • Keywords
    blood pressure measurement; blood vessels; band-type cuff; cuff unit; cuff-pressure controller; finger artery; instantaneous blood pressure; local pressurization; volume-compensation method; Arteries; Biomedical monitoring; Blood pressure; Fingers; Measurement units; Mechanical variables measurement; Pressure control; Pressure measurement; Servosystems; Volume measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Bioinformatics and Biomedical Engineering , 2009. ICBBE 2009. 3rd International Conference on
  • Conference_Location
    Beijing
  • Print_ISBN
    978-1-4244-2901-1
  • Electronic_ISBN
    978-1-4244-2902-8
  • Type

    conf

  • DOI
    10.1109/ICBBE.2009.5162194
  • Filename