• DocumentCode
    2497291
  • Title

    Algorithms for Correcting Velocity Vectors in Ultrasonic Particle Image Velocimetry

  • Author

    Lili Niu ; Hairong Zheng

  • fYear
    2009
  • fDate
    11-13 June 2009
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    We have recently developed a novel contrast-based ultrasonic particle imaging technique (Echo PIV) for multi-dimensional blood flow imaging. In the technique, the quality of vectors is important for quantitative blood flow imaging. Here we introduce an algorithm for vectors correction by filtering in frequency domain and applying continuity equation. The algorithm was tested in the Echo PIV imaging of rotational flow field and found to be effective in improving the quantitative blood flow imaging quality.
  • Keywords
    biomedical ultrasonics; blood flow measurement; flow visualisation; frequency-domain analysis; medical signal processing; ultrasonic imaging; continuity equation; contrast based ultrasonic PIV; echo PIV; frequency domain filtering; multidimensional blood flow imaging; particle image velocimetry; velocity vector correction algorithm; Bandwidth; Blood flow; Equations; Filtering algorithms; Frequency domain analysis; Ultrasonic imaging; Ultrasonic transducers; Vectors; Velocity control; Velocity measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Bioinformatics and Biomedical Engineering , 2009. ICBBE 2009. 3rd International Conference on
  • Conference_Location
    Beijing
  • Print_ISBN
    978-1-4244-2901-1
  • Electronic_ISBN
    978-1-4244-2902-8
  • Type

    conf

  • DOI
    10.1109/ICBBE.2009.5162299
  • Filename
    5162299