Title :
Laboratory in a box: Wearable sensors and its advantages for gait analysis
Author :
Najafi, Bijan ; Khan, Tahir ; Wrobel, James
Author_Institution :
Center for Lower Extremity Ambulatory Res., Rosalind Franklin Univ. of Med. & Sci. (RFUMS), Chicago, IL, USA
fDate :
Aug. 30 2011-Sept. 3 2011
Abstract :
Until recently, many gait studies explored potential gait alteration due to various disorders in the gait lab and using camera based systems and force platforms. However, these strategies may not replicate normal outdoor walking. Using this equipment, it is more difficult to measure the variability of walking which is important for maintaining balance and responding to different walking challenges. Additionally, subjects may mask their problem or exaggerate it when they are walking in a short walking distance offered by laboratory based-technology. This study overviews some of the key advantages of wearable technology compared to laboratory-based instrument. Additionally, it explored gait patterns over ample distance of walking compared to walking distance restricted to a gait laboratory environment. Walking patterns of ten healthy young subjects were examined using a wearable sensor technology in a random order over a distance of 7m, 14m, and 20m. Results suggest that participants walk significantly faster by increasing walking distance on average by 15% and 3% when walking distance was increased respectively from 7m to 14 and from 14m to 20m (p<;0.05). Interestingly despite a high test-retest reliability for averaged gait parameters (ICC>;0.89), the test-retest reliability for gait variability was only acceptable during 20m walking distance (ICC<;0.3 for 7m and 14m v. ICC=0.65 for 20m). Taken together, our findings indicate that for valid and reliable assessment of gait parameters, gait should be performed over ample walking distances. Body worn sensor technology facilitates assessing gait outside of a gait laboratory, over ample walking distance, different footwear condition, different walking surface, and in environment where mimics better true environment where the subject is active in.
Keywords :
biomedical equipment; biomedical measurement; biosensors; gait analysis; medical disorders; reliability; camera based systems; force platforms; gait laboratory environment; gait patterns; laboratory based-technology; laboratory-based instrument; medical disorders; walking patterns; wearable sensor technology; wearable sensors; Biological system modeling; Biomechanics; Diabetes; Laboratories; Legged locomotion; Reliability; Wearable sensors; Adult; Aged; Aging; Biomechanics; Equipment Design; Gait; Humans; Monitoring, Ambulatory; Movement; Risk Assessment; Time Factors; Walking;
Conference_Titel :
Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
Conference_Location :
Boston, MA
Print_ISBN :
978-1-4244-4121-1
Electronic_ISBN :
1557-170X
DOI :
10.1109/IEMBS.2011.6091605