DocumentCode
2501607
Title
Effect of Temperature on Counting Measurements in a Uranium Enrichment Monitor Based on a NaI(Tl) Spectrometer and Transmission Source
Author
Ianakiev, K.D. ; Alexandrov, B.S. ; Close, D.A. ; Dale, D.J. ; Goda, J.M. ; Hill, T.R. ; Marks, T. ; Moss, C.E. ; Nguyen, H.
Author_Institution
Los Alamos Nat. Lab., NM
Volume
1
fYear
2006
fDate
Oct. 29 2006-Nov. 1 2006
Firstpage
552
Lastpage
556
Abstract
We have studied the stability of a NaI(T1) scintillation detector and a commercial, general-purpose, digital, multichannel analyzer (MCA) used in uranium enrichment monitoring and have evaluated their combined sensitivity to temperature. The method used to calculate enrichment is very sensitive to any error that may be introduced by the instrument when measuring the intensity of the transmission peak corrected by the dead time (DT). An error as small as 0.1% in the attenuation of the transmission-peak count rate results in about a 3% error in the enrichment calculation. The demand for such an unusually high level of accuracy challenges the capabilities of an off-the-shelf instrument. The experimental data show that a variation of +/-10degC in the room temperature can cause a 0.4% change in the count rate of the transmission peak, even though the gain stabilization system holds the transmission peak at a specified channel. We have observed that the temperature changes the MCA´s pileup rejector efficiency and, thus, the DT correction. This is because the shapers in the fast and slow channels have different temperature characteristics. We have utilized a custom-designed detector preamplifier that stabilizes the preamplifier pulse shape over the working temperature range.
Keywords
pulse height analysers; radiation monitoring; scintillation counters; NaI(T1) scintillation detector; NaI(T1) spectrometer; counting measurements; detector preamplifier; digital multichannel analyzer; gain stabilization system; preamplifier pulse shape stabilization; transmission peak count rate; transmission source; uranium enrichment monitoring; Error correction; Instruments; Monitoring; Preamplifiers; Scintillation counters; Spectroscopy; Stability analysis; Temperature measurement; Temperature sensors; Time measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2006. IEEE
Conference_Location
San Diego, CA
ISSN
1095-7863
Print_ISBN
1-4244-0560-2
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2006.356218
Filename
4179057
Link To Document