DocumentCode
2501618
Title
Methods for speeding up radiated and conducted immunity tests
Author
Pommerenke, David
Author_Institution
Hewlett Packard, Roseville, CA, USA
Volume
2
fYear
2000
fDate
2000
Firstpage
587
Abstract
Different methods for reducing test time in immunity tests similar to IEC 61000-4-6 or IEC 61000-4-3 are compared. One option is to apply multiple signals at the same time. Hardware and software methods to create those signals are explained, limits and possible amplitude errors are discussed. Mode stirred chambers and methods which aim at substituting sinusoidal excitation by strong pulsed signals are not within the scope of this paper
Keywords
IEC standards; electromagnetic interference; measurement errors; measurement standards; EMI; IEC 61000 standards; IEC 61000-4-3; IEC 61000-4-6; amplitude errors; conducted immunity tests; hardware methods; multiple signals; radiated immunity tests; sinusoidal excitation; software methods; test time reduction; Cables; Delay; Hardware; IEC standards; Maintenance engineering; Polarization; RF signals; Radio frequency; Signal generators; Testing;
fLanguage
English
Publisher
ieee
Conference_Titel
Electromagnetic Compatibility, 2000. IEEE International Symposium on
Conference_Location
Washington, DC
Print_ISBN
0-7803-5677-2
Type
conf
DOI
10.1109/ISEMC.2000.874686
Filename
874686
Link To Document