DocumentCode
2501694
Title
Electromagnetic interference effect from power line noise in electrocardiograph signal using faraday cage
Author
Zakaria, Noor Amalina ; Sudirman, Rubita ; Jamaluddin, Mohd Najeb
fYear
2008
fDate
1-3 Dec. 2008
Firstpage
666
Lastpage
671
Abstract
The main objective of this study is to build a basic human size faraday cage. 1-channel ECG circuit inclusive of notch filter also has been built to test the effect of faraday cage. The outcome of ECG signals was compared when placed inside cage (inside building), outside faraday cage (inside building) and outside building. Then the study of relationship between faraday cage and the effect on power line noise in ECG system has been done. From the result of this project it can be seen that faraday cage can eliminate the 50 Hz power line noise in ECG signal.
Keywords
Faraday effect; electrocardiography; electromagnetic interference; medical signal processing; power cables; 1-channel ECG circuit; ECG signals; Faraday cage; electrocardiograph signal; electromagnetic interference effect; frequency 50 Hz; notch filter; power line noise; Electromagnetic interference; IEC; ECG signal; Faraday cage; electromagnetic interference; power line noise;