• DocumentCode
    2501694
  • Title

    Electromagnetic interference effect from power line noise in electrocardiograph signal using faraday cage

  • Author

    Zakaria, Noor Amalina ; Sudirman, Rubita ; Jamaluddin, Mohd Najeb

  • fYear
    2008
  • fDate
    1-3 Dec. 2008
  • Firstpage
    666
  • Lastpage
    671
  • Abstract
    The main objective of this study is to build a basic human size faraday cage. 1-channel ECG circuit inclusive of notch filter also has been built to test the effect of faraday cage. The outcome of ECG signals was compared when placed inside cage (inside building), outside faraday cage (inside building) and outside building. Then the study of relationship between faraday cage and the effect on power line noise in ECG system has been done. From the result of this project it can be seen that faraday cage can eliminate the 50 Hz power line noise in ECG signal.
  • Keywords
    Faraday effect; electrocardiography; electromagnetic interference; medical signal processing; power cables; 1-channel ECG circuit; ECG signals; Faraday cage; electrocardiograph signal; electromagnetic interference effect; frequency 50 Hz; notch filter; power line noise; Electromagnetic interference; IEC; ECG signal; Faraday cage; electromagnetic interference; power line noise;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Power and Energy Conference, 2008. PECon 2008. IEEE 2nd International
  • Conference_Location
    Johor Bahru
  • Print_ISBN
    978-1-4244-2404-7
  • Electronic_ISBN
    978-1-4244-2405-4
  • Type

    conf

  • DOI
    10.1109/PECON.2008.4762559
  • Filename
    4762559