DocumentCode
2507165
Title
Implementation and Evaluation of a 3D PET Single Scatter Simulation with TOF Modeling
Author
Werner, Matthew E. ; Surti, Suleman ; Karp, Joel S.
Author_Institution
Pennsylvania Univ., Philadelphia, PA
Volume
3
fYear
2006
fDate
Oct. 29 2006-Nov. 1 2006
Firstpage
1768
Lastpage
1773
Abstract
Accurate quantification in TOF listmode PET reconstruction requires proper modeling of the TOF dependence of scatter. We have extended our non-TOF single scatter simulation to provide this information. We have observed excellent agreement with experimental data and an improvement in the uniformity of scatter correction over scaled non-TOF scatter correction. The method is practical in a clinical setting with existing computational resources. It can also provide simplifications to the reconstruction processing pipeline.
Keywords
image reconstruction; medical image processing; positron emission tomography; time of flight spectra; 3D PET single scatter simulation; TOF listmode PET reconstruction; TOF scatter correction; positron emission tomography; reconstruction processing pipeline; time-of-flight scatter correction; Detectors; Electromagnetic scattering; Event detection; Image reconstruction; Nuclear and plasma sciences; Particle scattering; Photonic crystals; Positron emission tomography; Radiology; Telephony;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2006. IEEE
Conference_Location
San Diego, CA
ISSN
1095-7863
Print_ISBN
1-4244-0560-2
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2006.354238
Filename
4179351
Link To Document