Title :
Initial Development of a Dual-Modality SPECT-CT System for Dedicated Mammotomography
Author :
Madhav, Priti ; Crotty, Dominic J. ; McKinley, L. ; Tornai, Martin P.
Author_Institution :
Dept. of Radiol., Duke Univ. Med. Center, Durham, NC
fDate :
Oct. 29 2006-Nov. 1 2006
Abstract :
Dual-modality systems offer great promise in improving detection and evaluation of cancer through enhancement of the visual quality and quantitative accuracy of radionuclide imaging. A compact SPECT-CT system for dedicated 3D breast imaging is in development. The SPECT components include a 16times20cm2 CZT-based compact gamma camera with 2.5mm square pixels, attached to a goniometer with polar positioning capability. The CT component includes a heavily filtered W-target X-ray source producing a quasi-monochromatic cone beam and CsI(Tl) digital detector. Both systems are coupled to a common rotation stage and have a common field of view. The CT system has a stationary polar orientation, and is laterally offset from the center of rotation for imaging pendant uncompressed breasts that are larger than the detector´s field of view. The constraints and performance of this initial configuration are being investigated. Previous measurements with combined system components show that optimal placement is limited by physical constraints rather than signal cross-contamination. Results on a combined system show that emission projection images are contaminated by X-ray scatter photons resulting in the reconstructed emission images having a higher signal level. Emission contamination also increases noise in the transmission image resulting in reducing the SNR in reconstructed CT images. Having both modalities on a single gantry is expected to simplify data acquisition, SPECT-CT image registration, and necessary image corrections.
Keywords :
X-ray scattering; cameras; cancer; goniometers; image registration; mammography; radioactive sources; single photon emission computed tomography; 16 cm; 20 cm; 3D breast imaging; CZT-based compact gamma camera; CsI(Tl) digital detector; SNR; X-ray scatter photons; cancer detection; cancer evaluation; data acquisition; dedicated mammotomography; dual-modality SPECT-CT system; filtered W-target X-ray source; gantry; goniometer; image corrections; image registration; pendant uncompressed breasts; polar positioning; quasi-monochromatic cone beam; radionuclide imaging; signal cross-contamination; stationary polar orientation; visual quality; Breast; Cameras; Cancer detection; Computed tomography; Digital filters; Goniometers; Image reconstruction; Optical imaging; X-ray detection; X-ray imaging;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2006. IEEE
Conference_Location :
San Diego, CA
Print_ISBN :
1-4244-0560-2
Electronic_ISBN :
1095-7863
DOI :
10.1109/NSSMIC.2006.354392