Geometrical Repeatability and Motion Blur Analysis of a New Multi-projection X-ray Imaging System
Author
Chawla, Amarpreet S. ; Samei, Ehsan
Author_Institution
Duke Adv. Imaging Labs., Duke Univ., Durham, NC
Volume
5
fYear
2006
fDate
Oct. 29 2006-Nov. 1 2006
Firstpage
3170
Lastpage
3173
Abstract
We report on the reproducibility and tests for possible motion blur artifacts of a new high frame-rate prototype X-ray multi-projection system recently installed in the Radiology Department of Duke University. The system, which has a translating X-ray tube but a stationary detector, is capable of acquiring oblique-angled full field projection images along both the vertical and horizontal axes at variable speeds and acquisition frame rates. The angular span of the tube movement is +7.5deg to -7.5deg about the posterior-anterior (PA) orientation. To analyze the effect of possible blurring artifact due to motion of the X-ray tube, multi-angled projection images of an edge device were acquired. The modulation transfer function (MTF) determined from the PA projection acquired with and without the tube in motion was compared to investigate blur artifacts that may be introduced due to motion of the X-ray tube. Geometrical precision of the system was assessed at two tube speeds by evaluating the recorded position coordinates as well as edge positions in repeated acquisitions. Excellent geometrical consistency was found in the tube motion. The positional errors at tube speeds of 1 and 2 in/sec were found to be within 1%. The possible effect on the MTF due to motion of the tube was found to be inconsequential. The system was deemed ready for actual clinical trials with these acquisition parameters.
Keywords
X-ray imaging; diagnostic radiography; image restoration; medical image processing; 0.0254 m/s; 0.0508 m/s; Duke University; MTF; Radiology Department; acquisition parameters; modulation transfer function; motion blur analysis; multiprojection X-ray imaging system; oblique angle full field projection images; positional errors; system geometrical precision; translating X-ray tube; Image analysis; Image motion analysis; Motion analysis; Prototypes; Radiology; Reproducibility of results; System testing; X-ray detection; X-ray detectors; X-ray imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2006. IEEE
Conference_Location
San Diego, CA
ISSN
1095-7863
Print_ISBN
1-4244-0560-2
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2006.356548
Filename
4179705
Link To Document