Abstract :
Results are presented for a low-cost, ultra-high resolution gamma camera for small-animal SPECT and molecular imaging. The detector, known as Bazooka SPECT, employs a second-generation image intensifier which is directly coupled to a columnar CsI(Tl) scintillator. Operating in photon-counting mode, individual gamma-ray interactions are seen as clusters of signal, and significant improvement in spatial resolution is obtained by estimating the interaction position via Anger (centroid) estimation for individual clusters. Amplification of scintillator light prior to the optical path is advantageous compared to similar, low-light CCD gamma-cameras which solely apply gain in the CCD. With amplification from the image intensifier, the system is no longer limited by light loss from the optical path. This allows for a customizable optical system via lenses in a macro photography configuration and for the use of a low-cost, highspeed CCD. Experimental results show that Bazooka SPECT has an intrinsic resolution of approximately 150mum when the entire 25mm intensifier output screen is imaged onto the CCD. Better than 100mum resolution is attainable with less demagnification from the optical system. Initial indications are that this detector will serve as an attractive, inexpensive modular camera for high-resolution, multiple-pinhole, small-animal SPECT and molecular imaging.
Keywords :
charge-coupled devices; image intensifiers; scintillation counters; single photon emission computed tomography; Anger estimation; Bazooka SPECT; columnar CsI(Tl) scintillator; columnar scintillators; high resolution gamma camera; image intensifiers; macrophotography configuration; molecular imaging; single photon imaging; small animal SPECT; Cameras; Charge coupled devices; Detectors; High-resolution imaging; Image intensifiers; Image resolution; Molecular imaging; Optical imaging; Spatial resolution; Stimulated emission; EMCCD; SPECT imaging; columnar CsI(Tl); gamma-camera; image intensifier; microchannel plate (MCP); molecular imaging;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2006. IEEE
Conference_Location :
San Diego, CA
Print_ISBN :
1-4244-0560-2
Electronic_ISBN :
1095-7863
DOI :
10.1109/NSSMIC.2006.353763