    Temperature Dependence of Spectroscopic Performance of Thallium Bromide X- and Gamma-Ray Detectors

  • Author

    Onodera, Toshiyuki ; Hitomi, Keitaro ; Shoji, Tadayoshi

  • Author_Institution
    Dept. of Electron., Tohoku Inst. of Technol., Sendai
  • Volume
    6
  • fYear
    2006
  • fDate
    Oct. 29 2006-Nov. 1 2006
  • Firstpage
    3711
  • Lastpage
    3714
  • Abstract
    Thallium bromide (TlBr) is a compound semiconductor characterized with wide band-gap energy (2.68 eV), high atomic numbers (Tl: 81, Br: 35) and high density (7.56 g/cm3). Therefore, TlBr is a suitable material for fabrication of X- and gamma-ray detectors. In this study, TlBr detectors have been fabricated from the crystals grown by the traveling molten zone method. The detectors have thickness of approximately 0.4 mm, and have gold electrodes (3 mm in diameters) on both surfaces of the crystals. In order to evaluate the temperature dependence of the charge transport properties of the detectors, the mobility-lifetime products for electrons ((mutau)e) and holes ((mutau)h) in the detectors have been estimated by the Hecht analysis at various temperatures of the detectors (-40, 20, 0, 20 and 40 degC). The (mutau)h in the detectors was of the order of 10-5 cm2/V at -40 degC, and increased with increasing temperatures. The maximum value of the (mutau)h was obtained at 20 degC, and was about ten times larger than the value at -40 degC. On the other hand, the (mutau)e was approximately 1 times 10-3 cm2/V between -40 degC and 20 degC. The temperature dependence of spectral response of the detectors has been evaluated by acquiring energy spectra of gamma-rays from 241Am at -40, 20, 0, 20 and 40 degC. At -40 degC, the full-energy peak corresponding to 59.5 keV gamma-rays has been obtained from the detectors, but the energy resolution of the peak was degraded due to the hole tailing on the low-energy side of the peak. The energy resolution of the detectors gradually improved with increasing temperatures of the detectors. The best FWHM energy resolution was measured to be 5.4 keV (9.1 %) at 20 degC.
  • Keywords
    X-ray detection; carrier lifetime; crystal growth from melt; electron mobility; gamma-ray detection; semiconductor counters; thallium compounds; wide band gap semiconductors; -20 C; -40 C; 0 C; 20 C; 40 C; Hecht analysis; X-ray detector; charge transport properties; electron mobility-lifetime product; energy resolution; gamma-ray detector; high atomic number semiconductor; high density semiconductor; hole tailing; holes mobility-lifetime product; thallium bromide; traveling molten zone method; wide band-gap semiconductor; Crystalline materials; Crystals; Energy resolution; Fabrication; Gamma ray detection; Gamma ray detectors; Photonic band gap; Semiconductor materials; Spectroscopy; Temperature dependence;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2006. IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1095-7863
  • Print_ISBN
    1-4244-0560-2
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2006.353799
  • Filename
    4179841