DocumentCode :
2519081
Title :
THREE-POINT METHOD FOR FAST AND ROBUST FIELD MAPPING FOR EPI GEOMETRIC DISTORTION CORRECTION
Author :
Aksit, Pelin ; Derbyshire, John A. ; Prince, Jerry L.
Author_Institution :
GE Healthcare, Johns Hopkins Univ., Laurel, MD
fYear :
2007
fDate :
12-15 April 2007
Firstpage :
141
Lastpage :
144
Abstract :
Because of its superb temporal resolution, echo planar imaging (EPI) is widely used for functional magnetic resonance imaging (fMRI), diffusion tensor imaging (DTI), and monitoring dynamic processes. EPI suffers from geometric distortions caused by magnetic field inhomogeneities, and although these distortions can be corrected by field mapping, current methods require lengthy auxiliary image acquisitions and/or phase unwrapping. We present a 3-point method for field mapping that circumvents these problems and is robust to additive noise. Results demonstrate that the performance of this approach is markedly better than the standard two-point method and is comparable to methods requiring much longer auxiliary image acquisitions
Keywords :
biomedical MRI; image resolution; EPI geometric distortion correction; additive noise; diffusion tensor imaging; echo planar imaging; field mapping; functional magnetic resonance imaging; image acquisitions; monitoring dynamic processes; temporal resolution; three-point method; Diffusion tensor imaging; Distortion measurement; Laboratories; Magnetic resonance imaging; Medical services; Noise robustness; Nonuniform electric fields; Phase distortion; Phase measurement; Radio frequency;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Biomedical Imaging: From Nano to Macro, 2007. ISBI 2007. 4th IEEE International Symposium on
Conference_Location :
Arlington, VA
Print_ISBN :
1-4244-0672-2
Electronic_ISBN :
1-4244-0672-2
Type :
conf
DOI :
10.1109/ISBI.2007.356808
Filename :
4193242
Link To Document :
بازگشت