• DocumentCode
    2519867
  • Title

    Improvement of the linearity of fast digital oscilloscopes used in averaging mode

  • Author

    Lembeye, Y. ; Keradec, J.-P. ; Cauffet, G.

  • Author_Institution
    Lab. d´´Electrotech. de Grenobe, Domaine Univ., Saint Martin d´´Heres, France
  • fYear
    1993
  • fDate
    18-20 May 1993
  • Firstpage
    146
  • Lastpage
    150
  • Abstract
    Accurate measurement of signals with very fast variations between values separated by several orders of magnitude is difficult. When signals are periodic, a fast digital oscilloscope can be used in an averaging mode but the measurement accuracy is mainly limited by analog to digital conversion (ADC) linearity. Several methods for improving the conversion linearity are proposed. These methods are theoretically justified and experimentally validated. Several of these methods can be used simultaneously by a digital oscilloscope to provide, in an averaging mode, with a 8-b ADC, the linearity of a 11-b or 12-b ADC. Residual errors are caused mainly by drifts of ADC characteristics. The compatibility of the proposed methods is discussed
  • Keywords
    analogue-digital conversion; calibration; digital instrumentation; measurement errors; oscilloscopes; signal processing; signal processing equipment; ADC; averaging mode; fast digital oscilloscopes; linearity; residual errors; Filtering; Linearity; Low pass filters; Noise figure; Noise measurement; Oscilloscopes; Power electronics; Q measurement; Quantization; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1993. IMTC/93. Conference Record., IEEE
  • Conference_Location
    Irvine, CA
  • Print_ISBN
    0-7803-1229-5
  • Type

    conf

  • DOI
    10.1109/IMTC.1993.382662
  • Filename
    382662