Title :
Enhancement of sensitivity of coulometric detection usng a one-electrode system
Author :
Ikemoto, K. ; Kojima, K. ; Sassa, F. ; Yokokawa, M. ; Suzuki, H.
Author_Institution :
Grad. Sch. of Pure & Appl. Sci., Univ. of Tsukuba, Tsukuba, Japan
Abstract :
Coulometric analysis is effective in the analysis of an analyte in a liquid plug of a very small volume. However, in the conventional coulometric detection, it took time to collect analyte molecules dissolved in the solution. To solve this problem, we used a novel method to achieve significant enhancement of sensitivity by using a one-electrode system. As a model analyte, hydrogen peroxide was oxidized in a flow channel and silver was deposited simultaneously on another electrode in another flow channel. Coulometric detection of the deposited silver showed a step change of generated charge. As a result, measurement time was reduced significantly (less than 2 s), although conventional methods took more than 60 s to achieve the same sensitivity. This shorter measurement time also suppressed the influence of background noise, resulting in significant lowering of the detection limit.
Keywords :
channel flow; electrochemical analysis; electrodes; analyte molecules; coulometric detection; coulometric detection sensitivity enhancement; flow channel; hydrogen peroxide; one-electrode system; Charge measurement; Electrodes; Fluid flow measurement; Plugs; Resists; Silver; Time measurement; Coulometry; Flow channel; Liquid plug; Microfluidics; One-electrode system;
Conference_Titel :
Solid-State Sensors, Actuators and Microsystems Conference (TRANSDUCERS), 2011 16th International
Conference_Location :
Beijing
Print_ISBN :
978-1-4577-0157-3
DOI :
10.1109/TRANSDUCERS.2011.5969242