• DocumentCode
    25317
  • Title

    Determination of ultrasound hydrophone phase from Fourier-Hilbert transformed 1 to 40 MHz time delay spectrometry amplitude

  • Author

    Bloomfield, Philip E. ; Lewin, Peter A.

  • Author_Institution
    Sch. of Biomed. Eng., Sci., & Health Syst., Drexel Univ., Philadelphia, PA, USA
  • Volume
    61
  • Issue
    4
  • fYear
    2014
  • fDate
    Apr-14
  • Firstpage
    662
  • Lastpage
    672
  • Abstract
    A Fourier-Hilbert transform (FHT) algorithm procedure was developed to obtain the complex characteristics of different hydrophones which included the most widely used membrane and capsule types. The theory presented was experimentally verified using magnitude response of the hydrophone probes determined with the time-delay-spectrometry (TDS) technique. The TDS setup implemented was similar to one used by Wear et al. in 2011, with the frequency range being extended to 1 to 40 MHz to ensure consistency with recently published international standards covering the properties, calibration, and performance evaluation of ultrasound hydrophones. The theoretical and experimental limitations of FHT were examined and revealed that extrapolations of the available logarithmic (volts per pascal magnitude) frequency response were needed to carry out acceptable FHT calculations of the phase response. The complex-valued analytic signal Hilbert transform (HT) algorithm was less practical in developing an efficient minimum phase calculation protocol than the equivalent real-valued discrete convolution HT algorithm. Examples of measured transfer function and FHT calculated phase response spectra of representative membrane and capsule ultrasound hydrophones are presented and are shown to obey the simple rule for minimum phase presented by Heyser in 1969.
  • Keywords
    Hilbert transforms; discrete Fourier transforms; frequency response; hydrophones; ultrasonic transducers; FHT algorithm; FHT calculated phase response spectra; Fourier-Hilbert transformed time delay spectrometry amplitude; HT algorithm; TDS technique; capsule types; capsule ultrasound hydrophones; complex-valued analytic signal Hilbert transform algorithm; equivalent real-valued discrete convolution HT algorithm; frequency 1 MHz to 40 MHz; hydrophone probes; logarithmic frequency response; magnitude response; measured transfer function; minimum phase calculation protocol; performance evaluation; phase response; representative membrane; time-delay-spectrometry technique; ultrasound hydrophone phase determination; Extrapolation; Frequency measurement; Frequency response; Phase measurement; Sonar equipment; Transforms; Ultrasonic imaging;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2014.2954