  • Abstract
    We present the design and calibration of a spectroscopic Stokes vector polarimeter. The polarimeter can measure the full Stokes vector in the wavelength range 450 nm to 767 nm with 1 nm resolution and consists of a fiber based spectrophotometer, a white LED light source, two liquid crystal retarders, and one polarizer. Calibration of the system is achieved with a scheme that does not require knowledge of the polarizing elements orientation or retardation. Six intensity spectra are required to calculate the full spectrum Stokes vector. Error in the polarimeter is less than 5%. We report the Stokes vectors of light transmitted through several polarizing element and biological media.
  • Keywords
    bio-optics; light emitting diodes; light transmission; liquid crystals; optical fibres; optical retarders; polarimeters; spectrophotometers; biomedical applications; calibration; fiber based spectrophotometer; intensity spectra; light transmission; liquid crystal retarders; polarizer; spectroscopic Stokes vector polarimetry; white LED light source; Biomedical measurements; Calibration; Light emitting diodes; Light sources; Liquid crystals; Optical fiber polarization; Optical polarization; Polarimetry; Spectroscopy; Wavelength measurement; Spectroscopy; Stokes vector;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    LEOS Annual Meeting Conference Proceedings, 2009. LEOS '09. IEEE
  • Conference_Location
    Belek-Antalya
  • ISSN
    1092-8081
  • Print_ISBN
    978-1-4244-3680-4
  • Electronic_ISBN
    1092-8081
  • Type

    conf

  • DOI
    10.1109/LEOS.2009.5343183
  • Filename
    5343183