• DocumentCode
    2536769
  • Title

    Rapid and repeated bolus drug delivery enabled by high efficiency electrochemical bellows actuators

  • Author

    Sheybani, R. ; Gensler, H. ; Meng, E.

  • Author_Institution
    Biomed. Microsyst. Lab., Univ. of Southern California, Los Angeles, CA, USA
  • fYear
    2011
  • fDate
    5-9 June 2011
  • Firstpage
    490
  • Lastpage
    493
  • Abstract
    For the first time, rapid and repeatable delivery of precise drug boluses (up to 40 15-second boluses of 19.12 ± 0.73 μL cocaine within 50 minutes) is achieved with a high efficiency MEMS electrolysis actuator. Gas bubble recombination parameters and drug viscosity were investigated. Nafion®-coated electrodes provided superior pumping performance (linearity and repeatability) compared to uncoated electrodes.
  • Keywords
    bioMEMS; biomedical electrodes; drug delivery systems; electrochemical devices; electrostatic actuators; microfluidics; Nafion coated electrodes; drug bolus delivery; drug viscosity; gas bubble recombination parameters; high efficiency MEMS electrolysis actuator; high efficiency electrochemical bellow actuators; pumping performance; rapid bolus drug delivery; repeated bolus drug delivery; Actuators; Bellows; Drugs; Electrochemical processes; Electrodes; Gases; Surface treatment; Bolus drug delivery; Nafion® polymer coating; drug pump; electrochemical actuators;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Solid-State Sensors, Actuators and Microsystems Conference (TRANSDUCERS), 2011 16th International
  • Conference_Location
    Beijing
  • ISSN
    Pending
  • Print_ISBN
    978-1-4577-0157-3
  • Type

    conf

  • DOI
    10.1109/TRANSDUCERS.2011.5969616
  • Filename
    5969616