The sigma.7 haptic interface for MiroSurge: A new bi-manual surgical console
Author :
Tobergte, Andreas ; Helmer, Patrick ; Hagn, Ulrich ; Rouiller, Patrice ; Thielmann, Sophie ; Grange, Sébastien ; Albu-Schäffer, Alin ; Conti, François ; Hirzinger, Gerd
Author_Institution :
DLR - German Aerosp. Center, Inst. of Robot. & Mechatron., Wessling, Germany
fYear :
2011
fDate :
25-30 Sept. 2011
Firstpage :
3023
Lastpage :
3030
Abstract :
This paper presents the design and control of the sigma.7 haptic device and the new surgical console of the MiroSurge robotic system. The console and the haptic devices are designed with respect to requirements in minimally invasive robotic surgery. Dedicated left and right handed devices are integrated in an operator console in an ergonomic configuration. The height of the whole console is adjustable, allowing the surgeon seated and standed operation. Each of the devices is fully actuated in seven degrees of freedom (DoF). A parallel mechanism with 3 DoF actuates the translational motion and an attached wrist with 3 intersecting axis drives the rotations of the grasping unit. This advantageous design leads to inherently decoupled kinematics and dynamics. Cartesian forces are 20 N within the translational workspace, which is a sphere of about 120 mm diameter for each device. The rotational wrist of the device covers the whole workspace of the human hand and provides maximum torques of about 0.4 Nm. The grasping unit can display forces up to 8 N. An integrated force/torque sensor is used to increase the transparency of the devices by reducing inertia and friction. It is theoretically shown that the non-linear closed loop system behaves like a passive system and experimental results validate the approach. The sigma.7 haptic devices are designed by Force Dimension in cooperation with the German Aerospace Center (DLR). DLR designed the surgical console and integrated the haptic devices in the MiroSurge system.
Keywords :
force sensors; haptic interfaces; medical robotics; surgery; MiroSurge robotic system; bi-manual surgical console; force sensor; minimally invasive robotic surgery; nonlinear closed loop system; sigma.7 haptic interface; torque sensor; Force; Grasping; Robot sensing systems; Surgery; Torque; Wrist;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Intelligent Robots and Systems (IROS), 2011 IEEE/RSJ International Conference on
Conference_Location :
San Francisco, CA
ISSN :
2153-0858
Print_ISBN :
978-1-61284-454-1
Type :
conf
DOI :
10.1109/IROS.2011.6094433
Filename :
6094433
Link To Document :
بازگشت