DocumentCode :
2544108
Title :
Calibration of EJ309 liquid scintillator for neutron spectrometry
Author :
Iwanowska, J. ; Swiderski, L. ; Krakowski, Tracey ; Kozlowski, Tadeusz ; Moszynski, M. ; Pausch, G.
Author_Institution :
Nat. Centre for Nucl. Res., Otwock-Swierk, Poland
fYear :
2012
fDate :
Oct. 27 2012-Nov. 3 2012
Firstpage :
149
Lastpage :
152
Abstract :
The aim of the work is the measurement of the recoil proton response of EJ309 liquid scintillator to neutrons in relation to gamma rays. The scintillation response for heavy particles is a non-linear function of the particle energy. In our work, we measure this response for EJ309 on fast neutrons and plot the curve of light output as a function of deposited energy. The method of the measurement is based on fast neutron scattering on protons in liquid scintillator medium. A simple setup of two detectors: EJ309 and plastic BC408 are used to register the energy of scattered neutron. We applied time of flight (ToF) technique to improve the accuracy of the method.
Keywords :
calibration; liquid scintillation detectors; neutron spectrometers; solid scintillation detectors; EJ309 liquid scintillator; calibration; deposited energy; fast neutron scattering; fast neutrons; gamma rays; heavy particles; light curve output; neutron spectrometry; nonlinear function; particle energy; plastic BC408; recoil proton response; time-of-flight technique; calibration; fast neutrons; liquid scintillators; time of flight;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location :
Anaheim, CA
ISSN :
1082-3654
Print_ISBN :
978-1-4673-2028-3
Type :
conf
DOI :
10.1109/NSSMIC.2012.6551081
Filename :
6551081
Link To Document :