Title :
A neutron sensitive microchannel plate detector with cross delay line readout
Author :
Berry, K.D. ; Bilheux, Hassina Z. ; Crow, Lowell ; Diawara, Y. ; Feller, W. Bruce ; Iverson, Erik B. ; Martin, Andrew ; Robertson, James L.
Author_Institution :
Oak Ridge Nat. Lab., Oak Ridge, TN, USA
fDate :
Oct. 27 2012-Nov. 3 2012
Abstract :
Microchannel plates containing neutron absorbing elements such as boron and gadolinium in the bulk glass are used as the sensing element in high spatial resolution, high rate neutron imaging systems. In this paper we describe one such device, using both 10B and natural Gd, which employs cross delay line signal readout, with time-of-flight capability. This detector has a measured spatial resolution under 40 μm FWHM, thermal neutron efficiency of 19%, and has recorded rates in excess of 500 kHz. A physical and functional description is presented, followed by a discussion of measurements of detector performance and a brief survey of some practical applications.
Keywords :
delay lines; microchannel plates; position sensitive particle detectors; readout electronics; cross delay line readout; neutron absorbing element; neutron imaging system; neutron sensitive microchannel plate detector; spatial resolution; thermal neutron efficiency;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location :
Anaheim, CA
Print_ISBN :
978-1-4673-2028-3
DOI :
10.1109/NSSMIC.2012.6551091