DocumentCode
2545360
Title
SiPM cross-talk amplification due to scintillator crystal: Effects on timing performance
Author
Gola, Alberto ; Tarolli, Alessandro ; Piemonte, C.
Author_Institution
Centro per i Mater. e i Microsistemi, Fondazione Bruno Kessler (FBK-TRST), Povo di Trento, Italy
fYear
2012
fDate
Oct. 27 2012-Nov. 3 2012
Firstpage
421
Lastpage
423
Abstract
The Dark Count Rate of the SiPM plays an important role in limiting the timing performance of the device when used for the readout of the light coming from a scintillator. We have already shown that it is possible to substantially reduce its effect on the Coincidence Resolving Time with a suitable noise compensation method. In this work we describe a different mechanism that limits the timing performance of the SiPM and it can be observed clearly when the contribution of the DCR is cancelled by the noise compensation. The mechanism worsens the CRT at higher over-voltages, actually limiting the maximum bias that can be applied to the device, and it seems to be related to the amplification of the optical cross-talk of the device due to the presence of the scintillator crystal. In this work we describe the experimental evidence of this phenomenon, its relation to the CRT performance of the devices as well as possible solutions.
Keywords
silicon radiation detectors; solid scintillation detectors; SiPM cross-talk amplification; SiPM dark count rate; SiPM timing performance; coincidence resolving time; device CRT performance; device optical cross-talk amplification; device timing performance; light readout; noise compensation; noise compensation method; scintillator crystal;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location
Anaheim, CA
ISSN
1082-3654
Print_ISBN
978-1-4673-2028-3
Type
conf
DOI
10.1109/NSSMIC.2012.6551139