DocumentCode
2547838
Title
Iterative reconstruction of coded source neutron radiographs
Author
Santos-Villalobos, H.J. ; Bingham, P.R. ; Gregor, J.
Author_Institution
Oak Ridge Nat. Lab., Oak Ridge, TN, USA
fYear
2012
fDate
Oct. 27 2012-Nov. 3 2012
Firstpage
968
Lastpage
971
Abstract
Use of a coded source facilitates high-resolution neutron imaging but requires that the radiographic data be deconvolved. In this paper, we compare direct deconvolution with two different iterative algorithms, namely, one based on direct deconvolution embedded in an MLE-like framework and one based on a geometric model of the neutron beam and a least squares formulation of the inverse imaging problem. Simulation results show that the model-based reconstruction method is the suitable solution for high-resolution coded source systems and where the illumination of the neutron source is non-uniform.
Keywords
high energy physics instrumentation computing; neutron detection; neutron sources; radiography; MLE-like framework; coded source neutron radiographs; direct deconvolution; high-resolution coded source systems; high-resolution neutron imaging; inverse imaging problem; iterative algorithms; least squares formulation; model-based reconstruction method; neutron beam geometric model; neutron source illumination; radiographic data;
fLanguage
English