Title :
Development of a Proton Computed Tomography system for pre-clinical tests
Author :
Civinini, C. ; Brianzi, M. ; Bruzzi, Mara ; Bucciolini, M. ; Carpinelli, M. ; Cirrone, G.A.P. ; Cuttone, G. ; Presti, D. Lo ; Pallotta, S. ; Pugliatti, C. ; Randazzo, N. ; Romano, Francesco ; Scaringella, Monica ; Sipala, V. ; Stancampiano, C. ; Talamonti
Author_Institution :
Florence Div., INFN, Florence, Italy
fDate :
Oct. 27 2012-Nov. 3 2012
Abstract :
Proton Computed Tomography (pCT) is an imaging method under investigation with the aim to improve the accuracy of treatment planning and patient positioning in hadron therapy. A pCT system should be capable to measure tissue electron density with an accuracy better than 1% and with a spatial resolution better than 1mm. A first apparatus based on a silicon tracker and a YAG:Ce crystal calorimeter, with a 5×5 cm2 active area, has been already developed by the authors in the framework of the PRIMA (Proton Imaging) collaboration and tested with proton beams with promising results. A second prototype with larger area (5×20 cm2) and improved data acquisition system is being designed. The details about this new prototype will be presented in this work.
Keywords :
biological tissues; calorimeters; computerised tomography; data acquisition; electron density; radiation therapy; PRIMA; YAG:Ce crystal calorimeter; data acquisition system; hadron therapy; pCT system; patient positioning; pre-clinical tests; proton beams; proton computed tomography; proton imaging; silicon tracker; spatial resolution; tissue electron density; treatment planning;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location :
Anaheim, CA
Print_ISBN :
978-1-4673-2028-3
DOI :
10.1109/NSSMIC.2012.6551313