Abstract
Combined PET-MRI scanners start a new era in medical imaging. However the development of MRI compatible PET detector module is a challenging task. SiPM sensors are insensitive to magnetic field and constitute a promising solution. A drawback is the high dark current. A readout concept for SiPM based small animal PET detector module is presented in this paper. The results show that the readout of the SiPM is possible using only four ADC channels and the position map is comparable to the ideal solution. The detector modules based on the method are feasible solution for MRI compatible PET scanners.
Keywords
biomedical MRI; biomedical equipment; photomultipliers; positron emission tomography; readout electronics; ADC channels; MRI compatible PET scanners; MRI compatible small animal PET detector; SiPM based small animal PET detector module; SiPM sensors; dark current; magnetic field; medical imaging; position map; time over threshold based readout method;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location
Anaheim, CA
ISSN
1082-3654
Print_ISBN
978-1-4673-2028-3
Type
conf
DOI
10.1109/NSSMIC.2012.6551317
Filename
6551317
Link To Document