Title :
Timing resolution improvement of MPPC for TOF-PET imaging
Author :
Nagano, Takeshi ; Sato, Kiminori ; Ishida, A. ; Baba, Toshihiko ; Tsuchiya, R. ; Yamamoto, Koji
Author_Institution :
Solid State Div., Hamamatsu Photonics K.K., Hamamatsu, Japan
fDate :
Oct. 27 2012-Nov. 3 2012
Abstract :
Large detection area and good timing resolution of Silicon Photomultiplier (SiPM) based scintillation detectors is required for TOF-PET imaging. Achieving good timing requires output pulse uniformity on each microcell to be good; this depends on minimizing the gain, the recovery time, the trace impedance variation, etc. A double metal trace, a metal resistor and a new structure were applied to produce uniform characteristics on a single Multi-Pixel Photon Counter (MPPC) channel. The resulting time resolution has been improved to 140 ps for approximately 10 photons irradiation and an overvoltage of 2.5 V on a 6×4 mnl single channel MPPC test pattern. Furthermore afterpulse probability has been dramatically suppressed to a typical value of 3% by applying new silicon wafers and new process conditions.
Keywords :
photomultipliers; positron emission tomography; silicon radiation detectors; solid scintillation detectors; TOF-PET timing; double metal trace; impedance variation; metal resistor; multipixel photon counter channel; output pulse uniformity; recovery time; silicon photomultiplier based scintillation detectors; silicon wafers; single-channel MPPC test pattern; timing resolution; trace impedance variation;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location :
Anaheim, CA
Print_ISBN :
978-1-4673-2028-3
DOI :
10.1109/NSSMIC.2012.6551376