• DocumentCode
    2553794
  • Title

    Towards using a Monolithic Active Pixel Sensor for in-vivo beam monitoring of Intensity Modulated Radiotherapy

  • Author

    Page, Ryan F. ; Abbott, Natalie L. ; Davies, John ; Dyke, Emma L. ; Randles, Heather J. ; Velthuis, J.J. ; Fletcher, Sam ; Gregory, Shaun D. ; Hall, C. ; John, Anne-Marie ; Lawrence, H. ; Stevens, Paul H. ; Hugtenburg, Richard P. ; Tunbridge, Victoria

  • Author_Institution
    H.H. Wills Phys. Lab., Univ. of Bristol, Bristol, UK
  • fYear
    2012
  • fDate
    Oct. 27 2012-Nov. 3 2012
  • Firstpage
    2200
  • Lastpage
    2204
  • Abstract
    To enable in-vivo monitoring of Intensity Modulated Radiotherapy (IMRT) with minimal attenuation of the treatment field, an upstream camera system based on a Monolithic Active Pixel Sensor (MAPS) has been developed. This system has been used to precisely reconstruct the position of the Multileaf Collimators (MLC) that are used to shape the treatment field. The results show the position of a MLC is known with a precision of 6 μm with 10 seconds worth of data and 52 ± 4 μm for a single frame.
  • Keywords
    radiation therapy; semiconductor counters; Multileaf Collimators; in-vivo beam monitoring; intensity modulated radiotherapy; monolithic active pixel sensor; treatment field minimal attenuation; upstream camera system;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
  • Conference_Location
    Anaheim, CA
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-2028-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2012.6551501
  • Filename
    6551501