Title :
Fast cone-beam-based LOR reconstruction for 3-D PET
Author :
Il Jun Ahn ; Kye Young Jeong ; Woo Ryun Nam ; Ji Rye Kim ; Jong Beom Ra
Author_Institution :
Dept. of Electr. Eng., KAlST, Daejeon, South Korea
fDate :
Oct. 27 2012-Nov. 3 2012
Abstract :
Iterative image reconstruction algorithms have been widely studied because they can provide better image quality than analytic reconstruction algorithms. In recent years, to achieve both good image quality and high computational efficiency, the line-of-response (LOR) reconstruction algorithm based on iteration had been developed. However, this algorithm requires a large amount of computation, which limits its practical application. There were several attempts to solve this computation problem by using a GPU. Since intrinsic LORs are not ordered, however, these attempts were limited to a ray-driven approach which is not suitable for parallel processing on a GPU. In this paper, we propose a fast cone-beam-based LOR reconstruction algorithm on a GPU. Experimental results show that the proposed algorithm reduces the computation time by about 31 times compared with a conventional GPU-based LOR reconstruction algorithm.
Keywords :
graphics processing units; image reconstruction; medical image processing; positron emission tomography; 3-D PET; GPU-based LOR reconstruction algorithm; computation problem; fast cone-bearn-based LOR reconstruction algorithm; image quality; iterative image reconstruction algorithms; line-of-response reconstruction algorithm;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location :
Anaheim, CA
Print_ISBN :
978-1-4673-2028-3
DOI :
10.1109/NSSMIC.2012.6551779