Title :
Improvements for automating voltage calibrations using a 10-V Josephson array
Author :
Steiner, R.L. ; Astalos, R.J.
Author_Institution :
Nat. Inst. of Stand. & Technol. Gaithersburg, MD, USA
Abstract :
Three changes that were introduced to automate totally a voltage standard system based on a 10-V Josephson array are described. A commercial standard cell scanner controls switching for calibrating either Zener references or digital voltmeters, a programmable attenuator helps in obtaining voltage steps, and measurements of digital voltmeter noise help in verifying array stability. The array system operates completely unattended.<>
Keywords :
Josephson effect; Zener diodes; calibration; digital voltmeters; superconducting junction devices; voltage measurement; 10 V; 10-V Josephson array; Zener references; automation; calibration; digital voltmeters; noise; programmable attenuator; programmable mm-wave attenuator; stability; standard cell scanner; voltage standard; voltage steps; Calibration; Circuits; Measurement standards; NIST; Noise level; Noise measurement; Switches; Voltage; Voltmeters; Wiring;
Conference_Titel :
Precision Electromagnetic Measurements, 1990. CPEM '90 Digest., Conference on
Conference_Location :
Ottawa, Ontario, Canada
DOI :
10.1109/CPEM.1990.109943