• DocumentCode
    2561348
  • Title

    A refrigerated Dewar for the Josephson array voltage calibration system

  • Author

    Duncan, R.V.

  • Author_Institution
    Sandia Nat. Lab., Albuquerque, NM, USA
  • fYear
    1990
  • fDate
    11-14 June 1990
  • Firstpage
    110
  • Lastpage
    111
  • Abstract
    A refrigerated Dewar has been used successfully with the Josephson array voltage calibration system to reliquefy helium with no degradation of the calibration system´s performance. The design of the Dewar is detailed, and experimental data obtained using it are reported. The independence of the array voltage from temperature has been confirmed to three parts in 10/sup 8/ per degree Kelvin using this Dewar.<>
  • Keywords
    Josephson effect; calibration; liquid helium; low-temperature techniques; superconducting junction devices; voltage measurement; He; Josephson array voltage calibration; array voltage; refrigerated Dewar; Calibration; Coaxial components; Helium; Josephson junctions; Manufacturing; NIST; Neck; Refrigeration; Tail; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements, 1990. CPEM '90 Digest., Conference on
  • Conference_Location
    Ottawa, Ontario, Canada
  • Type

    conf

  • DOI
    10.1109/CPEM.1990.109947
  • Filename
    109947