DocumentCode :
2561364
Title :
Evaluation of light extraction from PET detector modules using gamma equivalent UV excitation
Author :
Jatekos, Balazs ; Kettinger, Adam O. ; Lorincz, Emoke ; Ujhelyi, Ferenc ; Erdei, Gabor
Author_Institution :
Dept. of Atomic Phys., Budapest Univ. of Technol. & Econ., Budapest, Hungary
fYear :
2012
fDate :
Oct. 27 2012-Nov. 3 2012
Firstpage :
3746
Lastpage :
3750
Abstract :
Monolithic scintillator-based PET detector modules offer depth of interaction capability and lower price relative to conventional pixelated crystal detectors. However, making such constructions poses several problems that affect light extraction efficiency and accuracy of position determination. In this work we examine potential detector geometries proposed to solve these problems. The geometries were tested by our novel measurement method that utilizes point-like excitation of the scintillator. With the applied methods we successfully increased the light extraction with a factor of two relative to a conventional monolithic PET detector. The measurements were compared to results of complete optical CAD models of the investigated arrangements. We concluded that our optical CAD tool can be used to further optimize the current constructions.
Keywords :
positron emission tomography; solid scintillation detectors; PET detector modules; complete optical CAD models; conventional monolithic PET detector; conventional pixelated crystal detectors; gamma equivalent UV excitation; interaction capability depth; light extraction efficiency; light extraction evaluation; monolithic scintillator-based PET detector modules; optical CAD tool; position determination accuracy; scintillator point-like excitation;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location :
Anaheim, CA
ISSN :
1082-3654
Print_ISBN :
978-1-4673-2028-3
Type :
conf
DOI :
10.1109/NSSMIC.2012.6551860
Filename :
6551860
Link To Document :