    Theoretical and numerical analysis of the single-ring OpenPET geometry for in-beam PET

  • Author

    Tashima, Hideaki ; Hirano, Yoshikuni ; Kinouchi, Shin-ichi ; Yoshida, Erika ; Ito, H. ; Yamaya, Taiga

  • Author_Institution
    Nat. Inst. of Radiol. Sci., Chiba, Japan
  • fYear
    2012
  • fDate
    Oct. 27 2012-Nov. 3 2012
  • Firstpage
    3871
  • Lastpage
    3873
  • Abstract
    We are developing the OpenPET, an open type PET geometry having an accessible and observable open space between detector rings which allows easy access to a patient during PET measurements. Our first generation OpenPET which we called dual-ring OpenPET consisted of two detector rings having a gap between them and it could provide an extended axial field of view (FOV). For applications such as in-beam PET to monitor dose distribution in situ during particle therapy, however, higher sensitivity is required rather than a wide FOV. As the next generation OpenPET, we proposed the single-ring OpenPET which was a geometry having an accessible and observable open space with higher sensitivity and a reduced number of detectors than the earlier geometry. In this paper, we provide a theoretical analysis of the sensitivity for the single-ring OpenPET and we compare it with the dual-ring OpenPET and a geometry in which the conventional PET is positioned at a slant angle against the patient bed to form an accessible open space, which we called slant PET. The theoretical analysis shows that the single-ring OpenPET has a sensitivity 1.2 times higher than the dual-ring OpenPET and 1.3 times higher than the slant PET when designed for about 200 detector blocks, each with a front area of 2500 mm2. In addition, numerical simulation using the Geant4 Monte Carlo toolkit is done to show the three-dimensional imaging performance that the single-ring OpenPET realized with the ellipsoidal rings of block detectors. The numerical simulation demonstrates that the depth-of-interaction detector can provide uniform resolution even when the detectors are arranged in an ellipsoidal ring.
  • Keywords
    biomedical equipment; dosimetry; geometry; numerical analysis; positron emission tomography; radiation therapy; Geant4 Monte Carlo toolkit; PET measurements; block detectors; depth-of-interaction detector; detector rings; dose distribution in situ; dual-ring OpenPET; ellipsoidal rings; extended axial field of view; in-beam PET; numerical simulation; open space; particle therapy; single-ring OpenPET geometry; slant PET; slant angle; theoretical analysis; three-dimensional imaging performance; wide FOV;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
  • Conference_Location
    Anaheim, CA
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-2028-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2012.6551887
  • Filename
    6551887