• DocumentCode
    2562966
  • Title

    Calorimeter with compensative absorber for measuring mu W level laser power

  • Author

    Suzuki, Y. ; Murata, A. ; Araragi, M. ; Inoue, T.

  • Author_Institution
    Yokogawa Electr. Corp., Tokyo, Japan
  • fYear
    1990
  • fDate
    11-14 June 1990
  • Firstpage
    278
  • Lastpage
    279
  • Abstract
    A calorimeter with a compensative absorber developed for measuring mu W-level laser power is described. With this calorimeter, laser power measurements can be made irrespective of pressure fluctuations in the surroundings. The calorimeter was found to be very effective in reducing the effects of ambient pressure fluctuations. The pressure stability has been improved by more than tenfold, and the standard deviation of the measured value is 0.01 approximately 0.6% for 10-mW approximately 1- mu W power levels. The systematic error of the measurement, which is represented by the standard deviation (1 sigma ), was evaluated as +or-0.1% for 10 mu W.<>
  • Keywords
    calorimeters; laser variables measurement; measurement errors; measurement standards; power measurement; ambient pressure fluctuations; calorimeter; compensative absorber; laser power measurements; measurement standard; pressure fluctuations; pressure stability; standard deviation; systematic error; Cooling; Electric variables measurement; Fluctuations; Laboratories; Measurement standards; Optical sensors; Power lasers; Power measurement; Pressure measurement; Temperature sensors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements, 1990. CPEM '90 Digest., Conference on
  • Conference_Location
    Ottawa, Ontario, Canada
  • Type

    conf

  • DOI
    10.1109/CPEM.1990.110022
  • Filename
    110022