DocumentCode :
2562992
Title :
RF-DC differences of thermal voltage converters arising from input connectors (voltage measurement)
Author :
Huang, D.X. ; Kinard, J.R. ; Rebuldela, G.
Author_Institution :
Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
fYear :
1990
fDate :
11-14 June 1990
Firstpage :
280
Lastpage :
281
Abstract :
The RF-DC differences of thermal voltage converters (TVCs) caused by the skin effect and transmission line effects of input structures of different length have been previously studied. Discrepancies have been found to exist between simple mathematical models and measured results for commonly used input connectors. A study of these discrepancies is reported. In particular, the change in RF-DC difference for a 0.5-V TVC resulting from the insertion of a 10-cm air line is evaluated.<>
Keywords :
convertors; skin effect; transmission lines; voltage measurement; 0.5 V; 10 cm; AC/DC conversion; RF-DC differences; air line; input connectors; mathematical models; skin effect; thermal voltage converters; transmission line effects; voltage measurement; Calibration; Connectors; Impedance; NIST; Power transmission lines; Radio frequency; Skin effect; Transmission line measurements; Transmission lines; Voltage measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements, 1990. CPEM '90 Digest., Conference on
Conference_Location :
Ottawa, Ontario, Canada
Type :
conf
DOI :
10.1109/CPEM.1990.110023
Filename :
110023
Link To Document :
بازگشت