DocumentCode
2566465
Title
A 60/spl mu/W 60 nV/Hz Readout Front-End for Portable Biopotential Acquisition Systems
Author
Yazicioglu, Refet Firat ; Merken, P. ; Puers, Robert ; Van Hoof, Chris
Author_Institution
IMEC, Leuven
fYear
2006
fDate
6-9 Feb. 2006
Firstpage
109
Lastpage
118
Abstract
A biopotential readout front-end can be configured to extract EEG, ECG, and EMG signals and draws 20muA from 3V. AC coupling of chopped amplifiers results in an input-referred noise of 60nV/radicHz and CMRR of 120dB at 1kHz. The immunity of the CMRR to electrode offset voltages is improved with an active input stage and 110dB CMRR is achieved at 100Hz with 50mV electrode offset
Keywords
biomedical electrodes; data acquisition; electrocardiography; electroencephalography; electromyography; medical signal processing; portable instruments; readout electronics; 1 kHz; 100 Hz; 50 mV; 60 muW; CMRR; ECG; EEG; EMG; electrode offset voltages; portable biopotential acquisition systems; readout front-end; Choppers; Electrocardiography; Electrodes; Electroencephalography; Electromyography; Frequency; Immune system; Interference; Power dissipation; Voltage;