• DocumentCode
    2566465
  • Title

    A 60/spl mu/W 60 nV/Hz Readout Front-End for Portable Biopotential Acquisition Systems

  • Author

    Yazicioglu, Refet Firat ; Merken, P. ; Puers, Robert ; Van Hoof, Chris

  • Author_Institution
    IMEC, Leuven
  • fYear
    2006
  • fDate
    6-9 Feb. 2006
  • Firstpage
    109
  • Lastpage
    118
  • Abstract
    A biopotential readout front-end can be configured to extract EEG, ECG, and EMG signals and draws 20muA from 3V. AC coupling of chopped amplifiers results in an input-referred noise of 60nV/radicHz and CMRR of 120dB at 1kHz. The immunity of the CMRR to electrode offset voltages is improved with an active input stage and 110dB CMRR is achieved at 100Hz with 50mV electrode offset
  • Keywords
    biomedical electrodes; data acquisition; electrocardiography; electroencephalography; electromyography; medical signal processing; portable instruments; readout electronics; 1 kHz; 100 Hz; 50 mV; 60 muW; CMRR; ECG; EEG; EMG; electrode offset voltages; portable biopotential acquisition systems; readout front-end; Choppers; Electrocardiography; Electrodes; Electroencephalography; Electromyography; Frequency; Immune system; Interference; Power dissipation; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Solid-State Circuits Conference, 2006. ISSCC 2006. Digest of Technical Papers. IEEE International
  • Conference_Location
    San Francisco, CA
  • ISSN
    0193-6530
  • Print_ISBN
    1-4244-0079-1
  • Type

    conf

  • DOI
    10.1109/ISSCC.2006.1696039
  • Filename
    1696039