DocumentCode :
2567101
Title :
Inclusion of Inter Crystal Scatter data in PET
Author :
Gillam, J.E. ; Solevi, P. ; Oliver, J.F. ; Rafecas, M.
fYear :
2012
fDate :
2-5 May 2012
Firstpage :
62
Lastpage :
65
Abstract :
In PET, as the spatial resolution of the measurement system is increased, multiple interactions of a single photon may be separately measured and such events are often removed from the data used in image reconstruction. While for some PET imaging tasks this effect is unimportant, for primate, brain and high-spatial-resolution imaging where sensitivity is important, such Inter Crystal Scattering (ICS) events may constitute a large fraction of the measured data. On-the-fly list-mode image reconstruction is generally required in order to use all the information provided by ICS. Simulated One-Pass List-mode image reconstruction is used in this investigation to study the inclusion of ICS information into the reconstructed image using data acquired with the AX-PET prototype. In almost all cases an increase in image quality was observed, particularly for larger features. A small resolution penalty was also observed.
Keywords :
image reconstruction; medical image processing; positron emission tomography; AX-PET prototype; ICS information; high spatial resolution imaging; intercrystal scatter data; intercrystal scattering events; measurement system spatial resolution; on the fly image reconstruction; simulated one pass list mode image reconstruction; single photon interactions; Crystals; Image reconstruction; Photonics; Positron emission tomography; Spatial resolution; Standards; Image Reconstruction; Inter Crystal Scatter; List-mode; Positron Emission Tomography;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Biomedical Imaging (ISBI), 2012 9th IEEE International Symposium on
Conference_Location :
Barcelona
ISSN :
1945-7928
Print_ISBN :
978-1-4577-1857-1
Type :
conf
DOI :
10.1109/ISBI.2012.6235484
Filename :
6235484
Link To Document :