DocumentCode
2575572
Title
Single echo acquistion of MR images using RF coil arrays
Author
Wright, S.M. ; McDougall, M.P. ; Brown, D.G.
Author_Institution
Dept. of Electr. Eng., Texas A&M Univ., College Station, TX, USA
Volume
2
fYear
2002
fDate
2002
Firstpage
1181
Abstract
Parallel imaging methods such as SMASH and SENSE reduce imaging time by using receiver coil sensitivity patterns to reduce requirements for gradient based image localization. This paper describes the first use of a coil array to completely eliminate gradient phase encoding. By using a 64 element planar array and a custom built 64 channel receiver system, 64×256 resolution images were constructed from a single line of k-space, demonstrating the ability to form complete images from a single echo acquisition. Together, the receiver and array are capable of generating complete images during successive echo acquisitions, potentially enabling extremely rapid frame rates.
Keywords
arrays; biomedical MRI; biomedical equipment; coils; data acquisition; image coding; image resolution; medical image processing; MR images; RF coil arrays; SENSE; SMASH; extremely rapid frame rates; imaging time reduction; magnetic resonance imaging; medical diagnostic imaging; single echo acquistion; single line k-space; successive echo acquisitions; Acceleration; Coils; Data acquisition; Image resolution; Magnetic resonance; Magnetic resonance imaging; Phased arrays; Preamplifiers; Radio frequency; Sensor arrays;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology, 2002. 24th Annual Conference and the Annual Fall Meeting of the Biomedical Engineering Society EMBS/BMES Conference, 2002. Proceedings of the Second Joint
ISSN
1094-687X
Print_ISBN
0-7803-7612-9
Type
conf
DOI
10.1109/IEMBS.2002.1106336
Filename
1106336
Link To Document