Title :
Single echo acquistion of MR images using RF coil arrays
Author :
Wright, S.M. ; McDougall, M.P. ; Brown, D.G.
Author_Institution :
Dept. of Electr. Eng., Texas A&M Univ., College Station, TX, USA
Abstract :
Parallel imaging methods such as SMASH and SENSE reduce imaging time by using receiver coil sensitivity patterns to reduce requirements for gradient based image localization. This paper describes the first use of a coil array to completely eliminate gradient phase encoding. By using a 64 element planar array and a custom built 64 channel receiver system, 64×256 resolution images were constructed from a single line of k-space, demonstrating the ability to form complete images from a single echo acquisition. Together, the receiver and array are capable of generating complete images during successive echo acquisitions, potentially enabling extremely rapid frame rates.
Keywords :
arrays; biomedical MRI; biomedical equipment; coils; data acquisition; image coding; image resolution; medical image processing; MR images; RF coil arrays; SENSE; SMASH; extremely rapid frame rates; imaging time reduction; magnetic resonance imaging; medical diagnostic imaging; single echo acquistion; single line k-space; successive echo acquisitions; Acceleration; Coils; Data acquisition; Image resolution; Magnetic resonance; Magnetic resonance imaging; Phased arrays; Preamplifiers; Radio frequency; Sensor arrays;
Conference_Titel :
Engineering in Medicine and Biology, 2002. 24th Annual Conference and the Annual Fall Meeting of the Biomedical Engineering Society EMBS/BMES Conference, 2002. Proceedings of the Second Joint
Print_ISBN :
0-7803-7612-9
DOI :
10.1109/IEMBS.2002.1106336